article · Journal of Engineering Research and Reports
The project is titled Design and Construction of Electronic Muscle Stimulator (EMS) with embedded time Display unit and Dual Source of Powering. The EMS is a type of electrotherapy device that functions by stimulating muscle contraction using electrical impulses to strengthen weak muscles, reduce swelling, aid in blood circulations that helps to heal wounds. These impulses are generated by the device and deliver through electrode pads over the area that requires stimulation. The impulses from EMS mimic the action potential (the stimulus required to contract the muscle) that naturally comes from central nervous system (CNS). Proteus, software simulation Library was used to simulate the components and to understand the behaviour of the output stage. After that, hardware components were assembled together that gave rise to this EMS apparatus. Designing of the device was completed with different modes tagged Press, Thump, Rub, and vibrate with frequencies 40Hz, 20Hz, 10Hz, and 3Hz respectively. The project was completed and human experiment was carried and result showed that decrease in impedances brings increase in current intensity.
This page summarises published work. The authoritative version sits with the publisher.
DOI: 10.9734/jerr/2023/v25i4899
Is something wrong with this record? Report it or request removal.
Discussion
Have you built on this work, tried to replicate it, or seen it applied in practice? Share what you know. Verified researchers and MARATTO™ domain experts can open a discussion, and any member can reply. Contributions are reviewed before they appear.
No discussion yet. Open the first thread.
New to MARATTO™? Create a free account.