article · Delta University Scientific Journal
The aim of this study is to provide technical data bout replacing cerrobend casting blocks by multileaf collimator at the surface dose for different field sides and depths. As part of the commissioning process for a commercially available MLC, percentage depth doses from the surface to a depth of 30 cm have been measured for a number of square fields utilizing 6 MV and 15 MV X-rays. Changes in SSD, the existence of a secondary blocking tray in the beam, and radiation transmission between the MLC leaves were among the variables that were considered. We recommend using multileaf collimator as shaping technique instead of Cerrobend blocks as its economic and less time consuming.
This page summarises published work. The authoritative version sits with the publisher.
DOI: 10.21608/dusj.2023.318657
Is something wrong with this record? Report it or request removal.
Discussion
Have you built on this work, tried to replicate it, or seen it applied in practice? Share what you know. Verified researchers and MARATTO™ domain experts can open a discussion, and any member can reply. Contributions are reviewed before they appear.
No discussion yet. Open the first thread.
New to MARATTO™? Create a free account.