article · Reports of Practical Oncology & Radiotherapy
Background: EPID quality assurance (QA) workflows remain largely manual and are difficult to standardize across treatment machines. This limits efficiency, increases operator dependency, and complicates consistent reporting and data sharing between systems. Materials and methods: We integrated Pylinac image analysis into QATrack+ and adopted an API-oriented publication model to enable structured export and interoperability. Results: EPID images from an Elekta Versa HD (6–18 MV) were analyzed for field size, flatness, symmetry, and MLC Picket Fence metrics, showing close agreement. Conclusions: This workflow supports standardized, automated, and interoperable EPID QA reporting.
This page summarises published work. The authoritative version sits with the publisher.
DOI: 10.5603/rpor.112213
Is something wrong with this record? Report it or request removal.
Discussion
Have you built on this work, tried to replicate it, or seen it applied in practice? Share what you know. Verified researchers and MARATTO™ domain experts can open a discussion, and any member can reply. Contributions are reviewed before they appear.
No discussion yet. Open the first thread.
New to MARATTO™? Create a free account.